Line 9: Line 9:
 Smaller pumps (say < 50 L/s) tend to be passively cooled.  Somewhat larger pumps can be air cooled but really large ones I think are only water cooled. Smaller pumps (say < 50 L/s) tend to be passively cooled.  Somewhat larger pumps can be air cooled but really large ones I think are only water cooled.
  
 +In addition to the inlet and outlet flange, most (all?) of them also have a vent port.  This is supposed to help in venting the chamber without venting through the foreline pump which could push contamination into the chamber.
  
 +{{:equipment:vacuum:balzers:tph_050:connector.jpg?300|}}
  
-====== Controllers ======+Above: male receptacle used on balzers turbo pumps
  
  
 +====== Pump controllers ======
  
-====== Cables ======+{{:equipment:vacuum:balzers:tcp121:front.jpg?300|}} 
 +{{:equipment:vacuum:balzers:tcp121:back.jpg?300|}}
  
 +Above: front and rear of TCP-121 controller
  
 +Many of the controllers can be used with various pumps but mind that the connector wiring can differ despite being the same connectors.  Some of the smaller controllers (ex: TCP-015) use a smaller power connector so look carefully when matching cables.
 +
 +====== TCP121 ======
 +
 +Switches:
 +  * S9 (left): standby (operate at 66% full speed)
 +  * S2 (middle): heating, turbo pump
 +  * S1 (right): power.  Reset by holding power off for 2 seconds
 +
 +Connectors:
 +  * X1 (bottom circular): 
 +  * X2 (middle circular): 
 +  * X3 (top circular): 
 +  * X4 (right): 
 +  * X5 (left): turbo pump
 +
 +System connectors (referenced in schematics but not part of the TCP121 itself):
 +  * X8: circular connector on the turbo pump, goes to X5
 +
 +Fuses:
 +^ Fuse ^ Rating ^ Purpose ^
 +| F1 | 2.5A | Drive electronics |
 +| F2 | 0.2A | Vent valve |
 +| F3 | 2.5A | Heater |
 +| F4 | 0.5A | Cooling |
 +
 +
 +====== Station controllers ======
 +
 +Above: TCP303
 +
 +There are also some accessories that allow monitoring pump coolant, controlling vent valves, and other things.
 +
 +
 +====== Cabling ======
 +
 +===== Mains power =====
 +
 +==== Small plug (12 pin) ====
 +
 +{{:equipment:vacuum:balzers:mains_cable12:0.jpg?150|}}
 +
 +Above: A left, B right, 1 bottom, 6 top
 +
 +For TCP-040, TCP-300, & TCP-310 [WIRING: TCP-270]:
 +^ Purpose ^ Color ^ Pin ^
 +| Ground | green or green-yellow | A6 |
 +| Neutral | white or brown | A2 |
 +| Hot | black or blue | B1 |
 +
 +
 +==== Large plug (16 pin) ====
 +
 +Mains to X4 [TCP121 op inst, fig 10]:
 +
 +^ Purpose ^ Pin ^ 120V wire color ^
 +| Ground (PE / protective earth) | A8 | green |
 +| Neutral (N) | A2 | white |
 +| Hot (L / live) | B2 | black |
 +| Jumper | A1-B1 | N/A |
 +
 +Confirmed some of these by comparing schematics (which is somewhat hard to read) to pinout by pressing power switch
 +
 +
 +===== Controller to pump =====
 +
 +==== TCP120/121 to TPH/TPU 060, 240, 330 ====
 +
 +X5 (controller) to X8 (pump) [TCP121 op inst, fig 22].  Verified on "3m PM 011 232-X" cable
 +
 +^ Signal ^ TPU/TPH pump ^ TCP controller ^
 +|  | A | A5 |
 +|  | B | A4 |
 +|  | C | A3 |
 +|  | D | A2 |
 +|  | E | A1 |
 +|  | F | B6 |
 +|  | G | B5 |
 +|  | H | B4 |
 +|  | J | B3 |
 +|  | K | B2 |
 +|  | L | B1 |
 +|  | M | A6 |
 +|  | N | A7 |
 +|  | P | B7 |
 +|  | R | A1 |
 +|  | S | B8 |
 +|  | T | A2 |
 +|  | U | A3 |
 +|  | V | A8 |
 +
 +==== PM 441018-X ====
 +
 +Took the connector off of it to make above, pinned out by JM
 +
 +| A | a5 |
 +| B | a4 |
 +| C | a3 |
 +| D | a2 |
 +| E | a1 |
 +| F | b6 |
 +| G | a5 |
 +| H | a4 |
 +| J | a3 |
 +| K | a2 |
 +| L | a1 |
 +| N | a6 |
 +| M | N/C |
 +| P | N/C |
 +| R | N/C |
 +| S | N/C |
 +| T | N/C |
 +| U | N/C |
 +| V | N/C |
  
 ====== History ====== ====== History ======
Line 31: Line 150:
   * Rebuild kits   * Rebuild kits
     * http://www.sisweb.com/vacuum/turbo-repair-kits.htm     * http://www.sisweb.com/vacuum/turbo-repair-kits.htm
 +  * WIRING: TCP-270: http://www.idealvac.com/files/images/Pfeiffer_TCP-270_versus_040_300_310_AC_Cable_Drawings.pdf
 +  * TCP120/121 Operating Instructions (TCP121 op inst)
  