equipment:vacuum:balzers:turbo
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
equipment:vacuum:balzers:turbo [2013/08/10 23:07] – mcmaster | equipment:vacuum:balzers:turbo [2013/10/20 14:59] (current) – external edit 127.0.0.1 | ||
---|---|---|---|
Line 25: | Line 25: | ||
Many of the controllers can be used with various pumps but mind that the connector wiring can differ despite being the same connectors. | Many of the controllers can be used with various pumps but mind that the connector wiring can differ despite being the same connectors. | ||
+ | ====== TCP121 ====== | ||
- | ====== Station controllers ====== | + | Switches: |
+ | * S9 (left): standby (operate at 66% full speed) | ||
+ | * S2 (middle): heating, turbo pump | ||
+ | * S1 (right): power. | ||
+ | Connectors: | ||
+ | * X1 (bottom circular): | ||
+ | * X2 (middle circular): | ||
+ | * X3 (top circular): | ||
+ | * X4 (right): | ||
+ | * X5 (left): turbo pump | ||
- | Above: | + | System connectors (referenced in schematics but not part of the TCP121 itself): |
+ | * X8: circular connector on the turbo pump, goes to X5 | ||
+ | |||
+ | Fuses: | ||
+ | ^ Fuse ^ Rating ^ Purpose ^ | ||
+ | | F1 | 2.5A | Drive electronics | | ||
+ | | F2 | 0.2A | Vent valve | | ||
+ | | F3 | 2.5A | Heater | | ||
+ | | F4 | 0.5A | Cooling | | ||
+ | |||
+ | |||
+ | ====== Station controllers ====== | ||
+ | |||
+ | Above: | ||
There are also some accessories that allow monitoring pump coolant, controlling vent valves, and other things. | There are also some accessories that allow monitoring pump coolant, controlling vent valves, and other things. | ||
Line 36: | Line 59: | ||
====== Cabling ====== | ====== Cabling ====== | ||
- | ===== Power ===== | + | ===== Mains power ===== |
- | ==== Small plug ==== | + | ==== Small plug (12 pin) ==== |
+ | |||
+ | {{: | ||
+ | |||
+ | Above: A left, B right, 1 bottom, 6 top | ||
For TCP-040, TCP-300, & TCP-310 [WIRING: TCP-270]: | For TCP-040, TCP-300, & TCP-310 [WIRING: TCP-270]: | ||
Line 46: | Line 73: | ||
| Hot | black or blue | B1 | | | Hot | black or blue | B1 | | ||
+ | |||
+ | ==== Large plug (16 pin) ==== | ||
+ | |||
+ | Mains to X4 [TCP121 op inst, fig 10]: | ||
+ | |||
+ | ^ Purpose ^ Pin ^ 120V wire color ^ | ||
+ | | Ground (PE / protective earth) | A8 | green | | ||
+ | | Neutral (N) | A2 | white | | ||
+ | | Hot (L / live) | B2 | black | | ||
+ | | Jumper | A1-B1 | N/A | | ||
+ | |||
+ | Confirmed some of these by comparing schematics (which is somewhat hard to read) to pinout by pressing power switch | ||
+ | |||
+ | |||
+ | ===== Controller to pump ===== | ||
+ | |||
+ | ==== TCP120/121 to TPH/TPU 060, 240, 330 ==== | ||
+ | |||
+ | X5 (controller) to X8 (pump) [TCP121 op inst, fig 22]. Verified on "3m PM 011 232-X" cable | ||
+ | |||
+ | ^ Signal ^ TPU/TPH pump ^ TCP controller ^ | ||
+ | | | A | A5 | | ||
+ | | | B | A4 | | ||
+ | | | C | A3 | | ||
+ | | | D | A2 | | ||
+ | | | E | A1 | | ||
+ | | | F | B6 | | ||
+ | | | G | B5 | | ||
+ | | | H | B4 | | ||
+ | | | J | B3 | | ||
+ | | | K | B2 | | ||
+ | | | L | B1 | | ||
+ | | | M | A6 | | ||
+ | | | N | A7 | | ||
+ | | | P | B7 | | ||
+ | | | R | A1 | | ||
+ | | | S | B8 | | ||
+ | | | T | A2 | | ||
+ | | | U | A3 | | ||
+ | | | V | A8 | | ||
+ | |||
+ | ==== PM 441018-X ==== | ||
+ | |||
+ | Took the connector off of it to make above, pinned out by JM | ||
+ | |||
+ | | A | a5 | | ||
+ | | B | a4 | | ||
+ | | C | a3 | | ||
+ | | D | a2 | | ||
+ | | E | a1 | | ||
+ | | F | b6 | | ||
+ | | G | a5 | | ||
+ | | H | a4 | | ||
+ | | J | a3 | | ||
+ | | K | a2 | | ||
+ | | L | a1 | | ||
+ | | N | a6 | | ||
+ | | M | N/C | | ||
+ | | P | N/C | | ||
+ | | R | N/C | | ||
+ | | S | N/C | | ||
+ | | T | N/C | | ||
+ | | U | N/C | | ||
+ | | V | N/C | | ||
====== History ====== | ====== History ====== | ||
Line 60: | Line 151: | ||
* http:// | * http:// | ||
* WIRING: TCP-270: http:// | * WIRING: TCP-270: http:// | ||
+ | * TCP120/121 Operating Instructions (TCP121 op inst) | ||
equipment/vacuum/balzers/turbo.1376176053.txt.gz · Last modified: 2013/10/20 14:59 (external edit)