| ====== TCP121 ====== | ====== TCP121 ====== | ||
| + | |||
| + | Switches: | ||
| + | * S9 (left): standby (operate at 66% full speed) | ||
| + | * S2 (middle): heating, turbo pump | ||
| + | * S1 (right): power. | ||
| Connectors: | Connectors: | ||
| Line 39: | Line 44: | ||
| Fuses: | Fuses: | ||
| ^ Fuse ^ Rating ^ Purpose ^ | ^ Fuse ^ Rating ^ Purpose ^ | ||
| - | | F1 | 2.5A | | | + | | F1 | 2.5A | Drive electronics |
| - | | F2 | 0.2A | | | + | | F2 | 0.2A | Vent valve | |
| - | | F3 | 2.5A | | | + | | F3 | 2.5A | Heater |
| - | | F4 | 0.5A | | | + | | F4 | 0.5A | Cooling |
| - | ====== Station controllers ====== | ||
| + | ====== Station controllers ====== | ||
| - | Above: | + | Above: |
| There are also some accessories that allow monitoring pump coolant, controlling vent valves, and other things. | There are also some accessories that allow monitoring pump coolant, controlling vent valves, and other things. | ||
| Line 70: | Line 75: | ||
| ==== Large plug (16 pin) ==== | ==== Large plug (16 pin) ==== | ||
| - | |||
| - | XXX: diagram is a little iffy, probably disassemble to confirm this | ||
| Mains to X4 [TCP121 op inst, fig 10]: | Mains to X4 [TCP121 op inst, fig 10]: | ||
| - | ^ Purpose ^ Pin ^ | + | ^ Purpose ^ Pin ^ 120V wire color ^ |
| - | | Ground | A8 | | + | | Ground |
| - | | Neutral | A2 | | + | | Neutral |
| - | | Hot | B2 | | + | | Hot (L / live) | B2 | black | |
| - | | Jumper | A1-B1 | | + | | Jumper | A1-B1 | N/A | |
| + | |||
| + | Confirmed some of these by comparing schematics (which is somewhat hard to read) to pinout by pressing power switch | ||
| Line 86: | Line 91: | ||
| ==== TCP120/121 to TPH/TPU 060, 240, 330 ==== | ==== TCP120/121 to TPH/TPU 060, 240, 330 ==== | ||
| - | X5 (controller) to X8 (pump) [TCP121 op inst, fig 22]: | + | X5 (controller) to X8 (pump) [TCP121 op inst, fig 22]. Verified on "3m PM 011 232-X" cable |
| ^ Signal ^ TPU/TPH pump ^ TCP controller ^ | ^ Signal ^ TPU/TPH pump ^ TCP controller ^ | ||
| - | | | S | B8 | | + | | | A | A5 | |
| - | | | V | A8 | | + | | | B | A4 | |
| - | | | P | B7 | | + | | | C | A3 | |
| - | | | N | A7 | | + | | | D | A2 | |
| + | | | E | A1 | | ||
| | | F | B6 | | | | F | B6 | | ||
| - | | | L | B1 | | + | | | G | B5 | |
| | | H | B4 | | | | H | B4 | | ||
| - | | | M | A6 | | ||
| | | J | B3 | | | | J | B3 | | ||
| | | K | B2 | | | | K | B2 | | ||
| - | | | G | B5 | | + | | | L | B1 | |
| + | | | M | A6 | | ||
| + | | | N | A7 | | ||
| + | | | P | B7 | | ||
| | | R | A1 | | | | R | A1 | | ||
| - | | | E | A1 | | + | | | S | B8 | |
| | | T | A2 | | | | T | A2 | | ||
| - | | | D | A2 | | ||
| | | U | A3 | | | | U | A3 | | ||
| - | | | C | A3 | | + | | | V | A8 | |
| - | | | B | A4 | | + | |
| - | | | A | A5 | | + | ==== PM 441018-X ==== |
| + | |||
| + | Took the connector off of it to make above, pinned out by JM | ||
| + | | A | a5 | | ||
| + | | B | a4 | | ||
| + | | C | a3 | | ||
| + | | D | a2 | | ||
| + | | E | a1 | | ||
| + | | F | b6 | | ||
| + | | G | a5 | | ||
| + | | H | a4 | | ||
| + | | J | a3 | | ||
| + | | K | a2 | | ||
| + | | L | a1 | | ||
| + | | N | a6 | | ||
| + | | M | N/C | | ||
| + | | P | N/C | | ||
| + | | R | N/C | | ||
| + | | S | N/C | | ||
| + | | T | N/C | | ||
| + | | U | N/C | | ||
| + | | V | N/C | | ||
| ====== History ====== | ====== History ====== | ||