Line 57: Line 57:
  
 {{gallery>:mcmaster:contamination:cleaning_enc28:07_acetone_shake.jpg}} {{gallery>:mcmaster:contamination:cleaning_enc28:07_acetone_shake.jpg}}
 +
 +Looks exactly the same after
 +
 +Conclusion: stuff is still stuck on
  
 ====== Step 8: acetone lens tissue wipe ====== ====== Step 8: acetone lens tissue wipe ======
Line 70: Line 74:
 {{gallery>:mcmaster:contamination:cleaning_enc28:10_sonic_acetone.jpg}} {{gallery>:mcmaster:contamination:cleaning_enc28:10_sonic_acetone.jpg}}
  
-====== Step 10: acetone lens tissue wipe ======+====== Step 11: acetone lens tissue wipe ======
  
 {{gallery>:mcmaster:contamination:cleaning_enc28:11_wipe.jpg}} {{gallery>:mcmaster:contamination:cleaning_enc28:11_wipe.jpg}}
Line 83: Line 87:
  
 Ultrasound was more effective at removing large particulates than high pressure acetone.   Ultrasound was more effective at removing large particulates than high pressure acetone.  
 +
 +Ultrasound does not effectively remove undissolved epoxy but it can be removed by wiping
 +
 +Tearing lens tissue for smaller allotments is a bad idea as it generated particulates.  The sheet should be used whole and moved around if it is to be used multiple times.
  