| + | Looks exactly the same after | ||
| + | |||
| + | Conclusion: stuff is still stuck on | ||
| ====== Step 8: acetone lens tissue wipe ====== | ====== Step 8: acetone lens tissue wipe ====== | ||
| Line 70: | Line 74: | ||
| {{gallery>: | {{gallery>: | ||
| - | ====== Step 10: acetone lens tissue wipe ====== | + | ====== Step 11: acetone lens tissue wipe ====== |
| {{gallery>: | {{gallery>: | ||
| Line 83: | Line 87: | ||
| Ultrasound was more effective at removing large particulates than high pressure acetone. | Ultrasound was more effective at removing large particulates than high pressure acetone. | ||
| + | |||
| + | Ultrasound does not effectively remove undissolved epoxy but it can be removed by wiping | ||
| + | |||
| + | Tearing lens tissue for smaller allotments is a bad idea as it generated particulates. | ||